翻译成英语:e跌落试验 按GBT48575-92《包装运输包装件-跌落试验方法》进行跌落试验在设备正常出厂包装条件下按GBT485717-2017《包装运输包装件基本试验-编制性能试验大纲的通用规则》等级Ⅰ的测试强度进行了面、棱、角部位的跌落试验。 实验条件:跌落高度600mm在经过试验后检验便携式吸痰器:1外包装箱无破损;2内部包材无破损;3内部样品整体结构完好无可视破损通电后设备开
E) Drop Test The drop test was conducted in accordance with GB/T4857.5-92 "Packaging Transport Packaging-Method of Drop Test". Under the normal factory packaging conditions, the drop test was performed on the surface, edge, and corner of the product at the testing intensity level I specified in GB/T4857.17-2017 "Basic testing procedures for packaging transport packaging-Compilation of performance test outlines and general rules". Experimental conditions: Drop height of 600mm. After the test, the portable suction device was inspected and met the following criteria:
- The outer packaging box was undamaged;
- The internal packaging materials were undamaged;
- The internal sample was intact with no visible damage. After being powered on, the device operated normally without any abnormalities. Conclusion: Under normal factory packaging conditions, within the claimed effective period and transportation conditions, this device can guarantee the integrity of the packaging. 3.8.4 Confirmation of Transportation and Storage Requirements
- Packaging The portable suction device is packaged in a specialized outer packaging box, which should include moisture-proof plastic bags and shock-absorbing materials. The box should also include a product certificate, product manual, and related accessories.
- Transportation The portable suction device is suitable for any conventional transportation method. During transportation, it should be protected from rain, snow, and should not be mixed with corrosive substances.
- Storage The portable suction device should be stored in a dry, ventilated, and non-corrosive storage room.
原文地址: https://www.cveoy.top/t/topic/b6pi 著作权归作者所有。请勿转载和采集!